The reliability of the strength knee extensor muscle using a knee muscle force measurement device

Document Type : Original article

Authors

1 Students' Research Office, MSc Student in Physiotherapy, School of Rehabilitation, Shahid Beheshti University of Medical Sciences, Tehran, Iran

2 Professor, PhD in Physiotherapy, School of Rehabilitation, Shahid Beheshti University of Medical Sciences, Tehran, Iran

3 Assistant Professor, PhD in Physiotherapy, School of Rehabilitation, Shahid Beheshti University of Medical Sciences, Tehran, Iran

4 MSc in Sport Physiotherapy

5 Students' Research Office, BSc Student in Physiotherapy, School of Rehabilitation, Shahid Beheshti University of Medical Sciences, Tehran, Iran

Abstract

Background and Aim: The purpose of the present study was to examine the reliability of an isometric knee muscle force measurement device.
Materials and Method: Thirty-four healthy non-athlete subjects (17 males and 17 females, aged between 19-25) participated in the study. The maximal isometric muscle strength of the knee extensor muscles were measured at different times and on different days using a newly designed knee muscle force measurement apparatus. On each occasion, participants performed three maximum voluntary contractions in each direction of knee extension movement. The interclass correlation coefficient (ICC) and standard deviation within subjects (Sw) were computed to evaluate the reliability and magnitude of measurement error between measurements.
Results: The results of the intra-rater, test-retest, and inter-rater reliabilities (ICCs ≥0.90, Sws ≤ 9.1 N) indicate that the knee muscle force measurements were highly repeatable and varied slightly between measurements. There were no statistically significant difference in knee muscle force measurements between times, days, and raters. Maximal isometric muscle strengths were significantly higher in males compared with that in females (p<0.000). Women's right and left knee muscle strengths were 52.28% of that in men in knee extension, respectively.
Conclusion: The isometric muscle force measurement device used in the present study appears to be a reliable and useful instrument for measuring the force of the knee extensors. It could also be suitable for the assessment of a physiotherapeutic or rehabilitation program.

Keywords

Main Subjects


  1. Dunn JC, Iversen MD: Interrater Reliability of knee Force obtained by Hand-held Dynamometer. Backpain.2003##
  2. Fransen M,Crosbie J,Edmonds J:  Isometric muscle force measurement for clinicians treating patients with osteoarthritis of the knee 2003,49(1):29-35.##
  3. Robert J,Goldman , Krik A, Reinbold,Z..Annette Iglarash,Lawrence M.et al:Phase I design and evaluation of an isometric muscle reeducation device for knee osteoarthritis rehabilitation. J Rehabil Res Dev. 2003;40(2):95-107.##
  4. Hunter SK,Thompson MW,Adams RD:Relationships among age-associated strength changes and physical activity level,limb dominance and muscle group in woman 2002,55(6):B264-73##
  5. Rantanen T,Avland K,Suminen H,et al.:Muscle strength as a predictor of onset of ADL dependence in people aged 75 years. Aging Clin Exp Res. 2002;14(3 Suppl):10-5.##
  6. Laufer Y,Ries JD,Leininger PM,Alon G:Quadriceps femoris muscle torques and fatigue enerated by neuromuscular electrical stimulation with three different waveforms. Phys Ther. 2001 Jul;81(7):1307-16.##
  7. Haas M: Statistical methodology for reliability studies. Journal of Manipulative and Physiological Therapeutics, 1991; 14(2): 119-128.##
  8. Domholdt E: Physical Therapy Research: Principle and Application. Philadelphia: WB. Saunders Company, 1993, pp 153-157.##
  9. Bland JM, Altman DG: Measurement error. British Medical Journal, 1996, 313: 744.##
  10. McNemar Q: Physiological Statistics. 3th edition, New York: John Wiley and Sons Inc., 1962, pp 145-158.##
  11. Hinderer SR, Hinderer KA: Quantitative methods of evaluation. In: J.A. Delisa and B.A. Gans (Eds.), Rehabilitation Medicine, Principle and Practice. Philadelphia: J.B. Lippincott Company, 1993, p 106.##
  12. Stokes M: Reliability and repeatability of methods of measuring muscle in physiotherapy. Physiotherapy Practice, 1985, 1:71-76.##
  13. Bohannon,R.W.,Bubela,D.J,Magasi,S.R. Gershon, R.C: Relative reliability of three objective test of limb muscle strength. Isokinet Exerc Sci. 2011;19(2).##
  14. Wang YC, Bohannon RW, Magasi SR, Hrynkiewicz B, Morales A, Gershon RC, Rymer Z. Testing of knee extension muscle strength: A comparison of two portable alternatives for the NIH toolbox study. Isokinetics and exercise science. 2011;19(3):163-8. ##
  15. Katoh M, Isozaki K, Sakanoue N, Miyahara T. Reliability of isometric knee extension muscle strength measurement using a hand-held dynamometer with a belt: a study of test-retest reliability in healthy elderly subjects. Journal of Physical Therapy Science. 2010;22(4):359-63. ##
  16. Dunn JC, Iversen MD. Interrater Reliability of Knee Muscle Forces Obtained by Hand-held Dynamometer from Elderly Subjects with Degenerative Back Pain. Journal of Geriatric Physical Therapy. 2003;26(3):23-9.##
Volume 5, Issue 1
March and April 2016
Pages 93-100
  • Receive Date: 26 May 2015
  • Revise Date: 01 September 2015
  • Accept Date: 14 October 2015
  • First Publish Date: 20 March 2016